MKF115 constant temperature and humidity box is a high-efficiency and energy-saving star product manufactured by Binder, Germany. MKF series BINDER constant temperature and humidity box is based on DIN standard and IEC standard, suitable for temperature at -40 degrees Celsius
MKF115 constant temperature and humidity box is a high-efficiency and energy-saving star product manufactured by Binder, Germany. MKF series BINDER constant temperature and humidity box is based on DIN standard and IEC standard, suitable for temperature between -40 degrees Celsius to 180 degrees Celsius, and humidity between 10 degrees Celsius. The test of all temperature, humidity and climate conditions between the percentage and 98%. The machine has a LCD touch screen, which is easy to operate and easy to edit. It can display the complete data related to the operating status of the system. If abnormal conditions occur during operation, The screen automatically displays the cause of the failure and provides troubleshooting methods. For the constant temperature and humidity performance test of the battery, the energy consumption of the MKF series constant temperature and humidity box is much lower than that of the ordinary climate box, which can effectively achieve the purpose of high efficiency, energy saving and cost reduction. , With its extensive program functions and uniform temperature conditions, it provides a good guarantee for the consistency of temperature and humidity in the battery test climate environment.
main advantage
• Homogeneous climatic conditions with APT.line™ preheating chamber
• Automatic water management system and waste water management system
• Extensive programming and data acquisition
• Responsive steam humidification technology
• Door heating
• High temperature control accuracy, good temperature consistency in the box
• Large heated viewing window
• Inner chamber made of stainless steel
• Efficient and energy saving
• BINDER Multi Management Software APT-COM™ 4 Basic version
technical parameter
describe | MKF115-400V | MKF115-480V-C |
item code | 9020-0379 | 9020-0357 |
internal volumeL | 115 | 228 |
Equipment net weight (unloaded)kg | 280 | 214 |
Maximum total loadkg | 60 | 60 |
Maximum load per shelfkg | 30 | 30 |
size--Internal dimensions | ||
widthmm | 600 | 600 |
highmm | 480 | 480 |
depthmm | 400 | 400 |
size- External dimensions (excluding attachments and connections) | ||
widthNet weight mm | 980 | 980 |
highNet weight mm | 1725 | 1725 |
depthNet weight mm | 865 | 865 |
wall distanceside mm | 200 | 200 |
wall distancerear mm | 300 | 300 |
Observation window widthmm | 288 | 288 |
Observation window heightmm | 222 | 222 |
Temperature performance data | ||
temperature rangeSpendC | -40SpendC-180SpendC | -40SpendC-180SpendC |
temperature deviationdepending on set value ± K | 0.1...1.5 | 0.1...1.5 |
temperature fluctuationdepending on set value ± K | 0.1...0.5 | 0.1...0.5 |
average heating rateAccording to IEC 60068-3-5K\/min | 5.5 | 5.5 |
cooldownFrom 180 degrees Celsius to -40 degrees Celsius min | 120 | 120 |
average cooling rateAccording to IEC 60068-3-5K\/min | 4.5 | 4.5 |
maximum thermal compensationat 25 degrees Celsius W | 2800 | 2800 |
Climate performance data | ||
temperature rangeSpendC | 10SpendC -95SpendC | 10SpendC -95SpendC |
temperature fluctuationdepending on set value ± K | 0.1...1.3 | 0.1...1.3 |
Humidity rangePercent Relative Humidity | 1098 | 1098 |
Humidity Fluctuationdepending on the set value | less than or equal to2,5±relative humidity | less than or equal to2,5±relative humidity |
Condensate condensation temperature rangeCelsius | 5094 | 5094 |
Electrical data | ||
rated powerkW | 4.8 | 4.8 |
equipment safety deviceA | 16 | 16 |
Power Requirements | AC380,50Hz,3P | AC480,50Hz,3P |
data related to the environment | ||
sound pressure leveldB(A) | 62 | 67 |
firmware | ||
Number of Shelves (Standard\/most) | 1\/4 | 1\/4 |