Why does the setting temperature of the high and low temperature box fail?

2022-01-18

Everyone knows about the high and low temperature box. It is to achieve the setting temperature, and there may be some faults in the whole process. Today, let's learn how to solve the temperature failure of the high and low temperature box?

6377076820958526308070210

1. If the set temperature cannot be reached, it may be that the humidification tube is broken. You can contact the manufacturer for after-sales service in advance or replace the new high and low temperature box immediately. If the temperature change does not meet the test temperature In the case of the value, you can check the household electrical system, and then check the faults one by one.

2. Naturally, it is also possible that the temperature camera is broken, so it will cause the machine to not run when the temperature is lower than the set specification temperature.

3. The time delay of the temperature acquisition system is too long or the heating tube is brittle and damaged, causing the specific temperature to be lower than the specified temperature, and the machinery and equipment have not been running normally or the temperature has not come up.

4. If the temperature rises immediately to cause temperature protection, then the control panel fails, and the operating instrument panel must be replaced.

5. If the temperature overshoot is strong, then the main parameters of the PID must be set first.

6. If the temperature of the equipment rises relatively slowly, check the air circulation system and check whether the adjustment baffle of the air circulation starts to work normally. If the temperature rises rapidly, it is necessary to check whether the motor of the air circulation is running. normal.

Since the high and low temperature box has an application period, the operation time is long, and there is no doubt that problems will occur over time. Usually, regular maintenance and maintenance can generally prevent most of them. If you encounter a relatively large bottleneck problem, you should contact the manufacturer immediately. .


Related News